Back to Search
Start Over
Effect of tibial component varus on wear in total knee arthroplasty
- Source :
- The Knee. 19:560-563
- Publication Year :
- 2012
- Publisher :
- Elsevier BV, 2012.
-
Abstract
- Introduction Malalignment can result in poor clinical outcomes and increased wear. However, component malalignment can occur even when overall limb mechanical axis is within the normal anatomic range. We studied the effect of component malalignment in the presence of acceptable knee alignment in knee arthroplasty. Methods Sixteen tibial inserts retrieved at revision surgery were laser-mapped to measure wear. Average implantation duration was 7.7 years (range, 1 to 13). Early (postprimary) and final (prerevision) radiographs were analyzed for overall alignment (limb, femoral and tibial components) and osteolysis. Results The tibial components were initially aligned in a mean of 1.3 ± 1.7° varus (range, –1.5 to 4.5°), which increased to 3.2 ± 2.9° (range, –2.0 to 8.0°) at the time of revision (p = 0.05). Tibial components initially placed in greater than 3° varus were associated with almost twice the volumetric penetration rate. Anatomic knee angles were 5.4 ± 0.9° valgus (range, 4.0 to 7.0°) in the post-primary radiographs and decreased in prerevision radiographs to 3.8 ± 2.6° (range, –1.0 to 7.5°), (p = 0.04). Discussion Tibial varus was associated with increased medial compartment wear and total wear, thus affecting osteolysis in addition to local destruction of the bearing surface. Varus malalignment as low as 3° may result in accelerated wear, even if overall limb alignment is nearly ideal. These results indicate that tibial component alignment is an important factor associated with tibial tray subsidence and polyethylene wear even when limb alignment is neutral.
- Subjects :
- Male
Reoperation
musculoskeletal diseases
medicine.medical_specialty
Osteolysis
Knee Joint
Surface Properties
medicine.medical_treatment
Radiography
Prosthesis Design
Bearing surface
medicine
Humans
Orthopedics and Sports Medicine
Femur
Tibia
Femoral component
Arthroplasty, Replacement, Knee
Aged
Retrospective Studies
Orthodontics
biology
business.industry
Bone Malalignment
Middle Aged
musculoskeletal system
biology.organism_classification
medicine.disease
Arthroplasty
Prosthesis Failure
Surgery
Valgus
Female
Polyethylenes
Knee Prosthesis
business
Follow-Up Studies
Subjects
Details
- ISSN :
- 09680160
- Volume :
- 19
- Database :
- OpenAIRE
- Journal :
- The Knee
- Accession number :
- edsair.doi.dedup.....f25d8e8c75c9de1a9c3bd6e060a9f2b3
- Full Text :
- https://doi.org/10.1016/j.knee.2011.11.003