Back to Search Start Over

DEPFET Macropixel Detectors for MIXS: Integration and Qualification of the Flight Detectors

Authors :
Danilo Miessner
Martin Hilchenbach
Peter Lechner
Rafael Strecker
B. Gunther
Matteo Porro
Florian Schopper
A. Stefanescu
T. Lauf
Alexander Bähr
P. Majewski
G. Lutz
Lothar Strüder
Ladislav Andricek
Ralf P. Richter
Jonas Reiffers
Johannes Treis
Heike Soltau
Gerhard Schaller
K. Hermenau
M. Schnecke
G. de Vita
Source :
IEEE Transactions on Nuclear Science. 59:2479-2486
Publication Year :
2012
Publisher :
Institute of Electrical and Electronics Engineers (IEEE), 2012.

Abstract

The Mercury imaging X-ray spectrometer (MIXS) on board of ESA's fifth cornerstone mission BepiColombo will be the first space instrument using DEpleted P-channel FET (DEPFET) based detectors. The MIXS spectrometer comprises two channels with identical focal plane detectors and is dedicated to energy resolved imaging of X-ray fluorescence from the mercurial surface. We report on the characterization, integration, and spectroscopic qualification of MIXS flight detectors. Detector chips were precharacterized at die level in order to select the best dies for integration and to do homogeneity and yield studies. Then, the detector chips were integrated to MIXS Detector Plane Arrays (DPAs), a complicated process due to the sophisticated mechanical structure, which allows the thermal decoupling of the detector from its readout and control chips. After integration, spectroscopic qualification measurements were done in order to analyze the detector performance and to prove the excellent spectroscopic performance of the DEPFET Macropixel detectors over a wide temperature range. The integration and spectroscopic qualification of all flight grade modules is now successfully completed.

Details

ISSN :
15581578 and 00189499
Volume :
59
Database :
OpenAIRE
Journal :
IEEE Transactions on Nuclear Science
Accession number :
edsair.doi.dedup.....f0cebf9472b5ab836583cdca9b56f62b
Full Text :
https://doi.org/10.1109/tns.2012.2211616