Analytic performance studies and clinical reproducibility of a real-time PCRassay for the detection of epidermal growth factor receptor gene mutations informalin-fixed paraffin-embedded tissue specimens of non-small cell lungcancer
- Source :
- BMC Cancer
- Publication Year :
- 2013
- Publisher :
- Springer Science and Business Media LLC, 2013.
-
Abstract
- Background Epidermal growth factor receptor (EGFR) gene mutations identify patients with non-small cell lung cancer (NSCLC) who have a high likelihood of benefiting from treatment with anti-EGFR tyrosine kinase inhibitors. Sanger sequencing is widely used for mutation detection but can be technically challenging, resulting in longer turn-around-time, with limited sensitivity for low levels of mutations. This manuscript details the technical performance verification studies and external clinical reproducibility studies of the cobas EGFR Mutation Test, a rapid multiplex real-time PCR assay designed to detect 41 mutations in exons 18, 19, 20 and 21. Methods The assay’s limit of detection was determined using 25 formalin-fixed paraffin-embedded tissue (FFPET)-derived and plasmid DNA blends. Assay performance for a panel of 201 specimens was compared against Sanger sequencing with resolution of discordant specimens by quantitative massively parallel pyrosequencing (MPP). Internal and external reproducibility was assessed using specimens tested in duplicate by different operators, using different reagent lots, instruments and at different sites. The effects on the performance of the cobas EGFR test of endogenous substances and nine therapeutic drugs were evaluated in ten FFPET specimens. Other tests included an evaluation of the effects of necrosis, micro-organisms and homologous DNA sequences on assay performance, and the inclusivity of the assay for less frequent mutations. Results A >95% hit rate was obtained in blends with >5% mutant alleles, as determined by MPP analysis, at a total DNA input of 150 ng. The overall percent agreement between Sanger sequencing and the cobas test was 96.7% (negative percent agreement 97.5%; positive percent agreement 95.8%). Assay repeatability was 98% when tested with two operators, instruments, and reagent lots. In the external reproducibility study, the agreement was > 99% across all sites, all operators and all reagent lots for 11/12 tumors tested. Test performance was not compromised by endogenous substances, therapeutic drugs, necrosis up to 85%, and common micro-organisms. All of the assessed less common mutations except one (exon 19 deletion mutation 2236_2248 > AGAC) were detected at a similar DNA input level as that for the corresponding predominant mutation. Conclusion The cobas EGFR Mutation Test is a sensitive, accurate, rapid, and reproducible assay.
- Subjects :
- Cancer Research
Lung Neoplasms
Companion diagnostics
Analytical validation
DNA Mutational Analysis
Gene mutation
Real-Time Polymerase Chain Reaction
medicine.disease_cause
symbols.namesake
Non-small cell lung cancer
Limit of Detection
Carcinoma, Non-Small-Cell Lung
Multiplex polymerase chain reaction
Molecular diagnostics
Genetics
medicine
Humans
Multiplex
Epidermal growth factor receptor
Sanger sequencing
Mutation
EGFR mutation testing
Paraffin Embedding
biology
Reproducibility of Results
Exons
Molecular biology
Reproducibility
ErbB Receptors
Real-time polymerase chain reaction
Technical Advance
Molecular Diagnostic Techniques
Oncology
symbols
biology.protein
Pyrosequencing
Multiplex Polymerase Chain Reaction
Subjects
Details
- ISSN :
- 14712407
- Volume :
- 13
- Database :
- OpenAIRE
- Journal :
- BMC Cancer
- Accession number :
- edsair.doi.dedup.....efcb37d23245037c0265473da28addf3
- Full Text :
- https://doi.org/10.1186/1471-2407-13-210