Back to Search
Start Over
Survey results of 3D-CRT and IMRT quality assurance practice
- Source :
- Journal of Applied Clinical Medical Physics
- Publication Year :
- 2019
-
Abstract
- Purpose To create a snapshot of common practices for 3D‐CRT and intensity‐modulated radiation therapy (IMRT) QA through a large‐scale survey and compare to TG‐218 recommendations. Methods A survey of 3D‐CRT and IMRT QA was constructed at and distributed by the IROC‐Houston QA center to all institutions monitored by IROC (n = 2,861). The first part of the survey asked about methods to check dose delivery for 3D‐CRT. The bulk of the survey focused on IMRT QA, inquiring about treatment modalities, standard tools used to verify planned dose, how assessment of agreement is calculated and the comparison criteria used, and the strategies taken if QA fails. Results The most common tools for dose verification were a 2D diode array (52.8%), point(s) measurement (39.0%), EPID (27.4%), and 2D ion chamber array (23.9%). When IMRT QA failed, the highest average rank strategy utilized was to remeasure with the same setup, which had an average position ranking of 1.1 with 90.4% of facilities employing this strategy. The second highest average ranked strategy was to move to a new calculation point and remeasure (54.9%); this had an average ranking of 2.1. Conclusion The survey provided a snapshot of the current state of dose verification for IMRT radiotherapy. The results showed variability in approaches and that work is still needed to unify and tighten criteria in the medical physics community, especially in reference to TG‐218's recommendations.
- Subjects :
- medicine.medical_specialty
Quality Assurance, Health Care
Computer science
Survey result
030218 nuclear medicine & medical imaging
03 medical and health sciences
0302 clinical medicine
Planned Dose
dose verification
medicine
Humans
Radiation Oncology Physics
Radiology, Nuclear Medicine and imaging
Medical physics
survey
Instrumentation
Dose delivery
Radiation
business.industry
Radiotherapy Planning, Computer-Assisted
patient specific QA
Radiotherapy Dosage
IMRT QA
Diode array
TG‐218
Ranking
Treatment modality
030220 oncology & carcinogenesis
Dose verification
Radiotherapy, Intensity-Modulated
Radiotherapy, Conformal
business
Quality assurance
Subjects
Details
- ISSN :
- 15269914
- Volume :
- 21
- Issue :
- 7
- Database :
- OpenAIRE
- Journal :
- Journal of applied clinical medical physics
- Accession number :