Evidence for Endotoxin Contamination in Plastic Na+-Heparin Blood Collection Tube Lots
- Source :
- Clinical Chemistry. 56:1483-1491
- Publication Year :
- 2010
- Publisher :
- Oxford University Press (OUP), 2010.
-
Abstract
- BACKGROUND Biomarker assays are often conducted on whole blood samples in the course of drug development studies. Because bacterial lipopolysaccharide (LPS) (endotoxin) contamination is known to cause spontaneous cytokine production by monocytes, contamination of blood collection tubes may interfere with biomarker assay results. METHODS Whole blood from healthy donors was collected into plastic or glass sodium (Na+)-heparin Vacutainer™ blood collection tubes and heparinized syringes. Samples were analyzed for phosphoprotein response, cytokine production, and RNA expression. Tubes were tested for endotoxin contamination by use of the limulus amoebocyte lysate assay. RESULTS Results of phospho-flow cytometry, branched DNA (bDNA), and ELISA assays indicated that a specific lot (#5339582) of plastic Na+-heparin Vacutainer tubes was highly contaminated with an endotoxinlike substance, and contamination was confirmed by the limulus amoebocyte lysate assay. Analysis of multiple-analyte panels revealed that analytes whose changed expression was predictive of LPS stimulation were increased when whole blood was incubated in contaminated tubes for 6 or 18 h. Two additional lots of plastic tubes tested had detectable amounts of endotoxin sufficient to strongly alter phospho-flow cytometry analyses, as determined by the fold change in phosphorylation of p38 mitogen-activated protein kinase in response to tumor necrosis factor α and LPS. In contrast, 3 lots of glass tubes had substantially lower levels of spontaneous blood activation. CONCLUSIONS Endotoxin contamination associated with tubes from 3 lots of a particular type of plastic Na+-heparin Vacutainer tube dramatically affected biomarker assay measurements. Prescreening these tubes is suggested before their use in clinical sample analysis.
- Subjects :
- Lipopolysaccharides
Time Factors
Lipopolysaccharide
Interleukin-1beta
Clinical Biochemistry
Enzyme-Linked Immunosorbent Assay
p38 Mitogen-Activated Protein Kinases
Flow cytometry
chemistry.chemical_compound
BDNA test
Humans
Medicine
RNA, Messenger
Chemokine CCL7
Phosphorylation
Chemokine CCL2
Whole blood
Blood Specimen Collection
biology
medicine.diagnostic_test
Heparin
Interleukin-6
Tumor Necrosis Factor-alpha
business.industry
Biochemistry (medical)
Anticoagulants
Contamination
Flow Cytometry
biology.organism_classification
Molecular biology
Endotoxins
Serum Amyloid P-Component
C-Reactive Protein
chemistry
Limulus
Immunology
Equipment Contamination
Blood Collection Tube
Vacutainer
business
Plastics
Biomarkers
Subjects
Details
- ISSN :
- 15308561 and 00099147
- Volume :
- 56
- Database :
- OpenAIRE
- Journal :
- Clinical Chemistry
- Accession number :
- edsair.doi.dedup.....ee736a60a535e227915cad721ea97ad1