Back to Search
Start Over
Combined LASER and femoral approach to remove a previous failure of Riata lead extraction
- Source :
- Clinical Case Reports
- Publication Year :
- 2017
- Publisher :
- Wiley, 2017.
-
Abstract
- Key Clinical Message Transvenous extraction of Riata lead could be a challenging procedure due to insulation failure and conductor cables’ externalization through the eroded outer insulation. In this case, a complete lead extraction was achieved through a combined LASER and femoral approach because of the massive conductors’ externalization.
- Subjects :
- insulation failure
business.industry
INSULATION FAILURE
Extraction (chemistry)
Case Report
Case Reports
General Medicine
030204 cardiovascular system & hematology
Laser
lead extraction
law.invention
03 medical and health sciences
Riata lead
0302 clinical medicine
law
Medicine
Conductor externalization
LASER
030212 general & internal medicine
Composite material
business
Electrical conductor
hybrid procedure
Lead extraction
Subjects
Details
- ISSN :
- 20500904
- Volume :
- 5
- Database :
- OpenAIRE
- Journal :
- Clinical Case Reports
- Accession number :
- edsair.doi.dedup.....ec951c2b664d2a78e66af7b803b72d94
- Full Text :
- https://doi.org/10.1002/ccr3.746