Comparison on the properties of bovine pericardium and porcine pericardium used as leaflet materials of transcatheter heart valve

Authors :
Wang Weidong
Hongping He
Zebin Wu
Ronghui Liu
Lizhen Wang
Li Zhao
Bin Li
Chang Jin
Source :
Artificial Organs. 46:427-438
Publication Year :
2021
Publisher :
Wiley, 2021.

Abstract

Background In order to obtain the smaller delivery diameter, porcine pericardium had been used as a substitute material of bovine pericardium for the leaflet materials of transcatheter heart valve (THV). However, the differences between them had not been fully studied. Therefore, this study compared the microstructure, biochemical and mechanical properties of two materials and hydrodynamics of THV made by the two materials in detail. Methods In this study, firstly, the microstructure of pericardium was analyzed by staining and scanning electron microscope; secondly, the biochemical properties of pericardium after different processes were compared by heat shrinkage temperature test, free amino and carboxyl concentration test, enzyme degradation test, subcutaneous implantation calcification analysis in rats; finally, the mechanical properties were evaluated by uniaxial tensile test before and after the pericardium being crimped, and then, the hydrodynamics of THV was studied according to the ISO5840 standard. Results Compared with bovine pericardium, after the same process, porcine pericardium showed a looser and tinier fiber bundle, a similar free carboxyl concentration, a lower resistance to enzyme degradation, a significantly lower calcification, bearing capacity and damage after being crimped, a better hydrodynamic and adaption with lower cardiac output and deformation of implantation position. Meanwhile the dehydration process of pericardium almost had preserved all the biochemical advantages of two materials. Conclusion In this study, porcine and bovine pericardium showed some significant differences in biochemical, mechanical properties and hydrodynamics. According to the results, it was presumed that the thinner porcine pericardium might be more suitable for THV of right heart system. Meanwhile, more attention should be taken for the calcification of THV made by the bovine pericardium.

Details

ISSN :
15251594 and 0160564X
Volume :
46
Database :
OpenAIRE
Journal :
Artificial Organs
Accession number :
edsair.doi.dedup.....ec2fadc2108d1acdcc68a8e2fa865e41
Full Text :
https://doi.org/10.1111/aor.14074