Improving the robustness of MOLLI T1 maps with a dedicated motion correction algorithm

Authors :
Gaspar Delso
Teresa M. de Caralt
Susanna Prat
Santi Sotes
José T. Ortiz-Pérez
Adelina Doltra
Marta Sitges
Martin A. Janich
Rosario J. Perea
Julián Vega
Laura Farré
Daniel Lorenzatti
Source :
Scientific Reports, Scientific Reports, Vol 11, Iss 1, Pp 1-9 (2021)
Publication Year :
2021
Publisher :
Springer Science and Business Media LLC, 2021.

Abstract

Myocardial tissue T1 constitutes a reliable indicator of several heart diseases related to extracellular changes (e.g. edema, fibrosis) as well as fat, iron and amyloid content. Magnetic resonance (MR) T1-mapping is typically achieved by pixel-wise exponential fitting of a series of inversion or saturation recovery measurements. Good anatomical alignment between these measurements is essential for accurate T1 estimation. Motion correction is recommended to improve alignment. However, in the case of inversion recovery sequences, this correction is compromised by the intrinsic contrast variation between frames. A model-based, non-rigid motion correction method for MOLLI series was implemented and validated on a large database of cardiac clinical cases (n = 186). The method relies on a dedicated similarity metric that accounts for the intensity changes caused by T1 magnetization relaxation. The results were compared to uncorrected series and to the standard motion correction included in the scanner. To automate the quantitative analysis of results, a custom data alignment metric was defined. Qualitative evaluation was performed on a subset of cases to confirm the validity of the new metric. Motion correction caused noticeable (i.e. > 5%) performance degradation in 12% of cases with the standard method, compared to 0.3% with the new dedicated method. The average alignment quality was 85% ± 9% with the default correction and 90% ± 7% with the new method. The results of the qualitative evaluation were found to correlate with the quantitative metric. In conclusion, a dedicated motion correction method for T1 mapping MOLLI series has been evaluated on a large database of clinical cardiac MR cases, confirming its increased robustness with respect to the standard method implemented in the scanner.

Details

ISSN :
20452322
Volume :
11
Database :
OpenAIRE
Journal :
Scientific Reports
Accession number :
edsair.doi.dedup.....ebe8e7715c6e74a010764d98f35bfbac
Full Text :
https://doi.org/10.1038/s41598-021-97841-z