Back to Search
Start Over
Electromagnetic interference to infusion pumps. Update2008 from GSM mobile phones
- Source :
- Scopus-Elsevier
-
Abstract
- Electromagnetic interference (EMI) to critical care medical devices has been reported by various groups. Previuos study demonstrated that infusion pumps are susceptible of false alarm buzzing and block of infusion, when exposed to various EMI sources. Aim of this paper is to investigate the changes in the risk of EMI from the estimates of our previous 2005 survey and to extend the EMI risk assessment to newer telecommunication products: DECT phones and WiFi terminals. With regards to GSM phones, compare to the results obtained in 2005, we observed a decrease in the rate of failure (from 58% to 30%). From our findings, the use of WiFi and DECT does not pose a real risk to infusion systems.
- Subjects :
- Engineering
business.industry
Digital Enhanced Cordless Telecommunications
Equipment Design
Risk Assessment
Electromagnetic interference
Equipment failure
Electromagnetic Fields
GSM
EMI
Electronic engineering
Equipment Failure
False alarm
Infusions, Intravenous
business
Risk assessment
Cell Phone
Infusion Pumps
Subjects
Details
- Database :
- OpenAIRE
- Journal :
- Scopus-Elsevier
- Accession number :