Lenstar® LS 900 vs Pentacam®-AXL: Comparative study of ocular biometric measurements and intraocular lens power calculation

Authors :
Pedro Alfaiate
João Carlos Sousa
Mónica Santos
Arminda Neves
J. M. Pereira
Henrique Aragão
Source :
European Journal of Ophthalmology. 28:645-651
Publication Year :
2018
Publisher :
SAGE Publications, 2018.

Abstract

Purpose: Comparison of biometric measurements and calculation of intraocular lens with a new biometer (Pentacam®-AXL, Oculus, Germany) and a reference biometer (Lenstar LS 900®, Haag-Streit AG, Switzerland), in order to assess the agreement between these two devices. Setting: Centro Hospitalar de Leiria, Portugal Materials and methods: Prospective, institutional study, in which measurements of axial length, anterior chamber depth from the corneal epithelium and endothelium to the anterior surface of the lens (anterior chamber depth ext and anterior chamber depth int), central corneal thickness and keratometry readings of the flattest and steepest meridians (K1 and K2) were obtained with the two systems. Intraocular lens calculation was also performed, using the Haigis, SRK/T and HofferQ formulas. Results: The study sample included 136 eyes of 79 patients. Of these, 42 were women and 37 were men. Statistically significant differences were found (p Conclusion: Axial length measurements obtained with the Pentacam® and Lenstar® appear to be interchangeable, while measurements of anterior chamber depth, K1 and K2, and central corneal thickness do not appear to be interchangeable between different devices. Statistically significant differences were found in the calculation of intraocular lenses in all formulas used.

Details

ISSN :
17246016 and 11206721
Volume :
28
Database :
OpenAIRE
Journal :
European Journal of Ophthalmology
Accession number :
edsair.doi.dedup.....ea13266b68995ec075687dc32e8664af
Full Text :
https://doi.org/10.1177/1120672118771844