Multi-Center Evaluation of the Automated Immunohematology Instrument, the ORTHO VISION Analyzer

Authors :
Ellen Klapper
Agnes Aysola
Katerina Pavenski
Elizabeth Krok
Connie Colavecchia
Leslie Wheeler
Rebecca Denham
Chelsea Hayes
Richard Brown
Source :
Laboratory Medicine. 48:29-38
Publication Year :
2017
Publisher :
Oxford University Press (OUP), 2017.

Abstract

BACKGROUND ORTHO VISION Analyzer (Vision), is an immunohematology instrument using ID-MT gel card technology with digital image processing. It has a continuous, random sample access with STAT priority processing. The efficiency and ease of operation of Vision was evaluated at 5 medical centers. METHODS De-identified patient samples were tested on the ORTHO ProVue Analyzer (ProVue) and repeated on the Vision mimicking the daily workload pattern. Turnaround times (TAT) were collected and compared. Operators rated key features of the analyzer on a scale of 1 to 5. RESULTS A total of 507 samples were tested on both instruments at the 5 trial sites. The mean TAT (SD) were 31.6 minutes (5.5) with Vision and 35.7 minutes (8.4) with ProVue, which renders a 12% reduction. Type and screens were performed on 381 samples; the mean TAT (SD) was 32.2 minutes (4.5) with Vision and 37.0 minutes (7.4) with ProVue. Antibody identification with eleven panel cells was performed on 134 samples on Vision; TAT (SD) was 43.2 minutes (8.3). The installation, training, configuration, maintenance and validation processes are all streamlined to provide a short implementation time. The average rating of main functions by the operators was 4.1 to 4.8. Opportunities for improvement, such as flexibility with editing QC results, maintenance schedule, and printing options were identified. The capabilities to perform serial dilutions, to accept pediatric tubes, and review results by e-Connectivity are enhancements over the ProVue. CONCLUSIONS Vision provides shorter TAT compared to ProVue. Every site described a positive experience using Vision.

Details

ISSN :
19437730 and 00075027
Volume :
48
Database :
OpenAIRE
Journal :
Laboratory Medicine
Accession number :
edsair.doi.dedup.....e8bc2beecb4d944dafa5f7e14c71ce83
Full Text :
https://doi.org/10.1093/labmed/lmw061