Neisseria gonorrhoeae False-Positive Result Obtained from a Pharyngeal Swab by Using the Roche cobas 4800 CT/NG Assay in New Zealand in 2012
- Publication Year :
- 2013
- Publisher :
- American Society for Microbiology, 2013.
-
Abstract
- The Roche cobas 4800 CT/NG assay is a commonly used commercial system for screening for Neisseria gonorrhoeae infection, and previous studies have shown the method to be highly sensitive and specific for urogenital samples. We present the first confirmed clinical N. gonorrhoeae false-positive result using the cobas 4800 NG assay, obtained from testing a pharyngeal swab sample and caused by cross-reaction with a commensal Neisseria strain.
- Subjects :
- Microbiology (medical)
Adult
Male
Recombination, Genetic
Pharyngeal swab
biology
Bacteriology
Chlamydia trachomatis
medicine.disease_cause
biology.organism_classification
Virology
Polymerase Chain Reaction
Neisseria gonorrhoeae
Microbiology
Highly sensitive
False-positive result
Gonorrhea