Quantum diamond microscopy with optimized magnetic field sensitivity and sub-ms temporal resolution
- Source :
- Journal of Applied Physics. 133
- Publication Year :
- 2023
- Publisher :
- AIP Publishing, 2023.
-
Abstract
- Quantum diamond magnetometers using lock-in detection have successfully detected weak bio-magnetic fields from neurons, a live mammalian muscle, and a live mouse heart. This opens up the possibility of quantum diamond magnetometers visualizing microscopic distributions of the bio-magnetic fields. Here, we demonstrate a lock-in-based wide-field quantum diamond microscopy, achieving a mean volume-normalized per pixel sensitivity of 43.9 $\mathrm{nTμm^{1.5}/Hz^{0.5}}$. We optimize the sensitivity by implementing a double resonance with hyperfine driving and magnetic field alignment along the $$ orientation of the diamond. Additionally, we show that sub-ms temporal resolution ($\sim$ 0.4 ms) can be achieved while keeping the per-pixel sensitivity at a few tens of nanotesla per second using quantum diamond microscopy. This lock-in-based diamond quantum microscopy could be a step forward in mapping functional activity in neuronal networks in micrometer spatial resolution.
Details
- ISSN :
- 10897550 and 00218979
- Volume :
- 133
- Database :
- OpenAIRE
- Journal :
- Journal of Applied Physics
- Accession number :
- edsair.doi.dedup.....e7d06b980e0251a89b02283d214778e5