Abstract

The traditional method of recording electrically evoked compound action potentials (ECAPs), as implemented in the MAESTRO clinical software (standard ART), requires manual adjustments during threshold determination through a specialist. The "FineGrain" research tool (FineGrain RT) uses a continuous stimulation paradigm combined with automatic ECAP threshold determination. The aim of this study was to compare the FineGrain RT with standard ART.ECAPs were recorded with standard ART and the FineGrain RT in paediatric cochlear implant recipients. Different stimulation rates were used for FineGrain ECAP recordings (40, 60, and 76 Hz).Thirteen children (6 - 19 years) participated in this study - nine were bilaterally and four unilaterally implanted, resulting in a total of twenty-two implanted ears.ECAP threshold determination success rates were similar between the two approaches (92% and 89%) and ECAP thresholds correlated well (r: 0.94,TheFineGrain research approach is a reliable replacement for standard ART in clinical practice.