Back to Search Start Over

Plan comparison of prostate stereotactic radiotherapy in spacer implant patients

Authors :
Keiji Kitatani
Shuichirou Miyazaki
Hiroshi Mayahara
Kazufusa Mizonobe
Tomonori Yabuuchi
Yuichirou Shiota
Hiroaki Akasaka
Risako Sakamoto
Aya Harada
Yuya Oki
Takayuki Hattori
Kazuyuki Uehara
Source :
Journal of Applied Clinical Medical Physics
Publication Year :
2021

Abstract

In prostate stereotactic body radiation therapy (SBRT), hydrogel spacers are increasingly used. This study aimed to perform a dosimetry comparison of treatment plans using CyberKnife (CK), commonly used for prostate SBRT, Helical TomoTherapy (HT), and TrueBeam (TB) in patients with hydrogel spacer implantations. The data of 20 patients who received hydrogel spacer implantation for prostate SBRT were retrospectively analyzed. The prescription dose was 36.25 Gy in five fractions to 95% of the planning target volume (PTV; D95). The conformity index (CI), gradient index (GI), homogeneity index (HI), and dose‐volume histogram (DVH) were analyzed for the three modalities, using the same PTV margins. The monitor unit (MU) and the beam‐on‐time (BOT) values were subsequently compared. The CI of TB (0.93 ± 0.02) was significantly superior to those of CK (0.82 ± 0.03, p