Back to Search
Start Over
Epinephrine auto-injector needle length : The impact of winter clothing
- Publication Year :
- 2020
- Publisher :
- Uppsala universitet, Pediatrisk inflammations- och metabolismforskning samt barnhälsa, 2020.
-
Abstract
- Background Epinephrine auto-injectors are expected to deliver the drug intramuscularly. Objective To study whether injection through clothing influences the frequency of subcutaneous and intraosseous/periosteal deposition of epinephrine. Methods Skin to muscle and skin to bone distances were measured for 303 children and adolescents and 99 adults. Distance was determined by ultrasound, with high or low pressure on the ultrasound probe. The risk/percentage of subcutaneous and intraosseous/periosteal injections was calculated using the lower and upper limits for the authority-approved length of EAI needles as provided by two high pressure EAI manufacturers and one low pressure EAI manufacturer. The addition winter clothing on the delivery of epinephrine was illustrated by comparing drug delivery fissue depth with no clothes. Furthermore, the riof non-intramuscular delivery for the shortest and longest approved needle length was calculated. Results When using EpipenJr® in children ® 0.1 mg had no risk of intraosseous/periosteal injection. However, the subcutaneous deposition risk increased from 94% and 28% to 100% and 99% with winter clothes. The risk of subcutaneous injection using EpipenJr® in the youngest children increased from 13% and 0% to 81% and 1% with winter clothes, and with Epipen® in adults from 45% and 17% to 60% and 38%. Emerade®, had a risk of subcutaneous injection in adults increasing from 14% and 10% to 28% and 21% adding winter clothes. Conclusion The risk of intraosseous/periosteal injections decreases and the risk of subcutaneous injection increases when injecting through winter clothes for all EAIs.
- Subjects :
- Periosteal
Epinephrine
Intraosseous
Clothing
03 medical and health sciences
Subcutaneous injection
0302 clinical medicine
Ultrasound probe
Auto-injector
parasitic diseases
Epinephrine Auto-Injector
Medicine
030212 general & internal medicine
Intramuscular
business.industry
Subcutaneous
Skin to muscle distance
Ultrasound
Public Health, Global Health, Social Medicine and Epidemiology
General Medicine
Auto-Injector
Folkhälsovetenskap, global hälsa, socialmedicin och epidemiologi
030228 respiratory system
High pressure
Anesthesia
Skin to bone distance