Back to Search
Start Over
Extracorporeal membrane oxygenation-induced fibrinolysis detected by rotational thromboelastometry and treated by oxygenator exchange
- Source :
- Perfusion. 34(4)
- Publication Year :
- 2019
-
Abstract
- Coagulopathy and bleeding is a frequent phenomenon in patients on extracorporeal membrane oxygenation. The cause may be multifactorial and it may change over time. We present a case when bleeding was caused by hyperfibrinolysis induced by oxygenator. The diagnosis was established by comparing thromboelastometry result from blood obtained before and after oxygenator. Hyperfibrinolysis and bleeding could be successfully treated merely by oxygenator exchange.
- Subjects :
- Change over time
Adult
Male
medicine.medical_treatment
030204 cardiovascular system & hematology
Oxygenators
03 medical and health sciences
0302 clinical medicine
Extracorporeal Membrane Oxygenation
Fibrinolysis
Coagulopathy
medicine
Extracorporeal membrane oxygenation
Humans
Radiology, Nuclear Medicine and imaging
In patient
Oxygenator
Advanced and Specialized Nursing
business.industry
General Medicine
medicine.disease
Hyperfibrinolysis
Thrombelastography
Thromboelastometry
030228 respiratory system
Anesthesia
Cardiology and Cardiovascular Medicine
business
Safety Research
Subjects
Details
- ISSN :
- 1477111X
- Volume :
- 34
- Issue :
- 4
- Database :
- OpenAIRE
- Journal :
- Perfusion
- Accession number :
- edsair.doi.dedup.....e4364fe9c486d12a73f8f439f019b0ea