Back to Search
Start Over
Shape Memory Tube Plug for Fine-control of Intraocular Pressure by Glaucoma Devices
- Source :
- ACS biomaterials scienceengineering. 6(7)
- Publication Year :
- 2021
-
Abstract
- As a standard approach to treating glaucoma, a silicone tube device is inserted into the eye to drain water and thereby lower the intraocular pressure (IOP). However, the fixed diameter of the tube often results in too much initial water drain. Suture-sheaving of the outer tube wall also leads to random IOP drops over time. In this study, self-expansion of the inner tube diameter was achieved by inserting a shape memory tube into the silicone tube. The difference in controlling small to large IOP drops before and after tube diameter expansion was demonstrated via computational modeling, a flow pumping system, and rabbit experiments.
- Subjects :
- Intraocular pressure
Materials science
Tube diameter
genetic structures
Biomedical Engineering
Glaucoma
law.invention
Biomaterials
Prosthesis Implantation
Tonometry, Ocular
law
medicine
Animals
Tube (fluid conveyance)
Spark plug
Glaucoma Drainage Implants
Silicone tube
Intraocular Pressure
Rabbit (nuclear engineering)
Shape-memory alloy
medicine.disease
eye diseases
sense organs
Rabbits
Biomedical engineering
Subjects
Details
- ISSN :
- 23739878
- Volume :
- 6
- Issue :
- 7
- Database :
- OpenAIRE
- Journal :
- ACS biomaterials scienceengineering
- Accession number :
- edsair.doi.dedup.....e3ead083c4e18e33e0ec48bd18d9a604