Back to Search Start Over

A survey of radiation dose to patients and operators during radiofrequency ablation using computed tomography

Authors :
C. A. Azlan
A. Saidatul
Ng Kh
Bjj Abdullah
Megat Amin M
Source :
Biomedical Imaging and Intervention Journal
Publication Year :
2010
Publisher :
Department of Biomedical Imaging, Faculty of Medicine, University of Malaya, Malaysia, 2010.

Abstract

Computed tomography (CT) fluoroscopy is able to give real time images to a physician undertaking minimally invasive procedures such as biopsies, percutaneous drainage, and radio frequency ablation (RFA). Both operators executing the procedure and patients too, are thus at risk of radiation exposure during a CT fluoroscopy. This study focuses on the radiation exposure present during a series of radio frequency ablation (RFA) procedures, and used Gafchromic film (Type XR-QA; International Specialty Products, USA) and thermoluminescent dosimeters (TLD-100H; Bicron, USA) to measure the radiation received by patients undergoing treatment, and also operators subject to scatter radiation. The voltage was held constant at 120 kVp and the current 70mA, with 5mm thickness. The duration of irradiation was between 150-638 seconds. Ultimately, from a sample of 30 liver that have undergone RFA, the study revealed that the operator received the highest dose at the hands, which was followed by the eyes and thyroid, while secondary staff dosage was moderately uniform across all parts of the body that were measured.

Details

Language :
English
ISSN :
18235530
Volume :
6
Issue :
1
Database :
OpenAIRE
Journal :
Biomedical Imaging and Intervention Journal
Accession number :
edsair.doi.dedup.....e385a0bba6331609ceeefbf14f2c298a