Performance of a cylindrical diode array for use in a 1.5 T MR-linac
- Source :
- Physics in medicine and biology, 61(3), N80-N89. IOP Publishing Ltd., Physics in Medicine and Biology, 61(3), N80. IOP Publishing Ltd.
- Publication Year :
- 2016
-
Abstract
- At the UMC Utrecht, a linear accelerator with integrated magnetic resonance imaging (MRI) has been developed, the MR-linac. Patient-specific quality assurance (QA) of treatment plans for MRI-based image guided radiotherapy requires QA equipment compatible with this 1.5 T magnetic field. The purpose of this study was to examine the performance characteristics of the ArcCHECK-MR in a transverse 1.5 T magnetic field. To this end, the short-term reproducibility, dose linearity, dose rate dependence, field size dependence, dose per pulse dependence and inter-diode dose response variation of the ArcCHECK-MR diode array were evaluated on a conventional linac and on the MR-linac. The ArcCHECK-MR diode array performed well for all tests on both linacs, no significant differences in performance characteristics were observed. Differences in the maximum dose deviations between both linacs were less than 1.5%. Therefore, we conclude that the ArcCHECK-MR can be used in a transverse 1.5 T magnetic field.
- Subjects :
- ArcCHECK-MR
MR-linac
Linear particle accelerator
030218 nuclear medicine & medical imaging
law.invention
03 medical and health sciences
0302 clinical medicine
Optics
Nuclear magnetic resonance
law
medicine
Journal Article
Radiology, Nuclear Medicine and imaging
ArcCHECK
Radiometry
radiotherapy
Physics
Medicine(all)
Reproducibility
MRI-accelerator
medicine.diagnostic_test
Radiological and Ultrasound Technology
Pulse (signal processing)
business.industry
Reproducibility of Results
patient specific QA
Magnetic resonance imaging
Particle accelerator
plan verification
Magnetic Resonance Imaging
Magnetic field
Transverse plane
Magnetic Fields
Radiology Nuclear Medicine and imaging
030220 oncology & carcinogenesis
Particle Accelerators
business
Quality assurance
Subjects