Back to Search Start Over

Ruptures of the device landing zone in patients undergoing transcatheter aortic valve implantation: an analysis of TAVI Karlsruhe (TAVIK) patients

Authors :
Jan Gerhardus
Armin Luik
Herbert Posival
Martin Heimeshoff
Peter Bramlage
Tim Süselbeck
Claus Schmitt
Rainer Wondraschek
Holger Schröfel
Gerhard Schymik
Source :
Clinical Research in Cardiology. 103:912-920
Publication Year :
2014
Publisher :
Springer Science and Business Media LLC, 2014.

Abstract

Aortic rupture of the device landing zone is a rare complication of transcatheter aortic valve implantation (TAVI) and it is associated with significant mortality.This study reports on the experience of a single-center in a case series of more than 1,000 implants. We explored patient and procedural characteristics aiming at identifying variables that increase the risk for aortic root rupture.Among a total of 1,000 TAVI procedures, six patients (0.6 %) had a rupture of the device landing zone. Five of these patients received the balloon-expandable Edwards SAPIEN valve (5/813; 0.62 %) of which four had a supraannular and one a subannular rupture. One patient received the self-expanding Medtronic CoreValve (1/199; 0.5 %; p = n.s. vs. SAPIEN) and had an annular rupture. Factors that were associated with aortic rupture were: (1) the relative size of the valve compared with the aortic annulus and its geometric form; (2) the need for post-dilation of the new valve because of paravalvular leakage; and (3) the location and severity of calcification. We determined, to avoid aortic rupture, caution may be necessary in the presence of the following conditions: (1) flat sinuses of Valsalva and severe calcifications of either the body or the free edge of the aortic cusps (supraannular rupture); (2) an ellipsoid annulus and bulky calcifications on either the base of the cusps or the rim of the annulus (annular rupture); and (3) a narrow left-ventricular outflow tract (LVOT) and bulky calcification of the LVOT (subannular rupture). After considering these precautions, we observed no case of aortic root rupture in the following 600 cases.The data indicate that to reduce the frequency of aortic rupture, a careful pre-procedural planning appears essential to avoid this serious and potentially deleterious complication.

Details

ISSN :
18610692 and 18610684
Volume :
103
Database :
OpenAIRE
Journal :
Clinical Research in Cardiology
Accession number :
edsair.doi.dedup.....e0d97ca3184fc0865ac93e591be2c40a
Full Text :
https://doi.org/10.1007/s00392-014-0729-8