Evaluation and verification of the QFix EncompassTM couch insert for intracranial stereotactic radiosurgery
- Source :
- Journal of Applied Clinical Medical Physics
- Publication Year :
- 2018
- Publisher :
- Wiley, 2018.
-
Abstract
- The QFix EncompassTM stereotactic radiosurgery (SRS) immobilization system consists of a thermoplastic mask that attaches to the couch insert to immobilize patients treated with intracranial SRS. This study evaluates the dosimetric impact and verifies a vendor provided treatment planning system (TPS) model in the Eclipse TPS. A thermoplastic mask was constructed for a Lucy 3D phantom, and was scanned with and without the EncompassTM system. Attenuation measurements were performed in the Lucy phantom with and without the insert using a pinpoint ion chamber for energies of 6xFFF, 10xFFF and 6X, with three field sizes (2 × 2, 4 × 4, and 6 × 6 cm2). The measurements were compared to two sets of calculations. The first set utilized the vendor provided Encompass TPS model (EncompassTPS), which consists of two structures: the Encompass and Encompass base structure. Three HU values for the Encompass (200, 300, 400) and Encompass Base (−600, −500, −400) structures were evaluated. The second set of calculations consists of the Encompass insert included in the external body contour (EncompassEXT) for dose calculation. The average measured percent attenuation in the posterior region of the insert ranged from 3.4%–3.8% for the 6xFFF beam, 2.9%–3.4% for the 10xFFF, and 3.3%–3.6% for the 6X beam. The maximum attenuation occurred at the region where the mask attaches to the insert, where attenuation up to 17% was measured for a 6xFFF beam. The difference between measured and calculated attenuation with either the EncompassEXT or EncompassTPS approach was within 0.5%. HU values in the EncompassTPS model that provided the best agreement with measurement was 400 for the Encompass structure and −400 for the Encompass base structure. Significant attenuation was observed at the area where the mask attaches to the insert. Larger differences can be observed when using few static beams compared to rotational treatment techniques.
- Subjects :
- treatment planning
Materials science
Dose calculation
medicine.medical_treatment
Posterior region
87.56.Da
dose calculation
Radiosurgery
Imaging phantom
030218 nuclear medicine & medical imaging
03 medical and health sciences
0302 clinical medicine
medicine
Radiation Oncology Physics
Humans
Radiology, Nuclear Medicine and imaging
87.53Ly
87.53Dq
Instrumentation
Insert (composites)
Radiation
Phantoms, Imaging
Radiotherapy Planning, Computer-Assisted
Attenuation
Radiotherapy Dosage
intracranial SRS
030220 oncology & carcinogenesis
immobilization
Ionization chamber
Radiotherapy, Intensity-Modulated
Beam (structure)
couch model
Biomedical engineering
Subjects
Details
- ISSN :
- 15269914
- Volume :
- 19
- Database :
- OpenAIRE
- Journal :
- Journal of Applied Clinical Medical Physics
- Accession number :
- edsair.doi.dedup.....dfc1d6120589c6274e3ec8f858e65f5a