Intraobserver Repeatability and Interobserver Reproducibility of Ellipsoid Zone Measurements in Retinitis Pigmentosa
- Source :
- Translational Vision Science & Technology
- Publication Year :
- 2018
- Publisher :
- The Association for Research in Vision and Ophthalmology, 2018.
-
Abstract
- Purpose To examine repeatability and reproducibility of ellipsoid zone (EZ) width measurements in patients with retinitis pigmentosa (RP) using a longitudinal reflectivity profile (LRP) analysis. Methods We examined Bioptigen optical coherence tomography (OCT) scans from 48 subjects with RP or Usher syndrome. Nominal scan lengths were 6, 7, or 10 mm, and the lateral scale of each scan was calculated using axial length measurements. LRPs were generated from OCT line scans, and the peak corresponding to EZ was manually identified using ImageJ. The locations at which the EZ peak disappeared were used to calculate EZ width. Each scan was analyzed twice by each of two observers, who were masked to their previous measurements and those of the other observer. Results On average, horizontal width (HW) was significantly greater than vertical width (VW), and there was high interocular symmetry for both HW and VW. We observed excellent intraobserver repeatability with intraclass correlation coefficients (ICCs) ranging from 0.996 to 0.998 for HW and VW measurements. Interobserver reproducibility was also excellent for both HW (ICC = 0.989; 95% confidence interval [CI] = 0.983-0.995) and VW (ICC = 0.991; 95% CI = 0.985-0.996), with no significant bias observed between observers. Conclusions EZ width can be measured using LRPs with excellent repeatability and reproducibility. Our observation of greater HW than VW is consistent with previous observations in RP, though the reason for this anisotropy remains unclear. Translational Relevance We describe repeatability and reproducibility of a method for measuring EZ width in patients with RP or Usher syndrome. This approach could facilitate measurement of retinal band thickness and/or intensity.
- Subjects :
- Physics
Reproducibility
retina
optical coherence tomography
medicine.diagnostic_test
business.industry
Intraclass correlation
Biomedical Engineering
Interobserver reproducibility
Repeatability
Articles
Reflectivity
Ellipsoid
03 medical and health sciences
Ophthalmology
0302 clinical medicine
Optical coherence tomography
030221 ophthalmology & optometry
medicine
In patient
ellipsoid zone
Nuclear medicine
business
030217 neurology & neurosurgery
Subjects
Details
- Language :
- English
- ISSN :
- 21642591
- Volume :
- 7
- Issue :
- 3
- Database :
- OpenAIRE
- Journal :
- Translational Vision Science & Technology
- Accession number :
- edsair.doi.dedup.....dfae4dc15877bebc5133cc740847dca2