Back to Search Start Over

Portable detection of serum HER-2 in breast cancer by a pressure-based platform

Authors :
Tianhai Ji
Dan Liu
Chaoyong James Yang
Qingyuan Lin
Haiyan Zheng
Xinyi Wu
Wensheng Yang
Qian Tao
Source :
Analytical and bioanalytical chemistry. 410(28)
Publication Year :
2018

Abstract

A high serum HER-2 extracellular domain (sHER-2 ECD) level has a reverse association with tumor behaviors. In this study, a portable platform for the disease biomarker sHER-2 ECD detection has been established using a pressure-based bioassay. The pressure bioassay consists of a monoclonal antibody immobilized on an eight-well strip, the analyte HER-2, and another monoclonal antibody labeled with the Pt nanoparticles (PtNPs), which have the catalytic ability to decompose H2O2 into H2O and O2(g). The increased pressure due to O2(g) generation is measured by a hand-held pressure meter. A total of 34 serum samples were collected to validate the performance of the pressure bioassay. The results showed that the pressure bioassay platform of HER-2 had a dynamic range from 2 to 50 ng/mL with a limit of detection (LOD) of 2 ng/mL, which was consistent with the ELISA result. In the real serum samples, there was a significant correlation between sHER-2 ECD level and several clinicopathological parameters, especially tissue HER-2 status. Furthermore, the sHER-2 ECD level was found to decrease after targeted therapy in a patient with tHER-2 positive. Overall, this bioassay can facilitate breast cancer diagnosis and prognosis in clinical scenarios and resource-limited areas.

Details

ISSN :
16182650
Volume :
410
Issue :
28
Database :
OpenAIRE
Journal :
Analytical and bioanalytical chemistry
Accession number :
edsair.doi.dedup.....dec066da74cb322522ad495648da9cd2