In the present paper, for the first time, the integrated three-electrode screen-printed sensor with in situ plated bismuth film carbon working electrode was applied for the ultratrace determination of thallium(I) (Tl(I)). Under optimized conditions extremely low limits of detection were reached, 8.47 × 10−10 and 6.71 × 10−12 mol L−1 for the deposition times of 60 s and 300 s, respectively. The influences of foreign metal ions and surfactants on the voltammetric signal of thallium in natural samples were minimized using 1 × 10−5 mol L−1 EDTA and Amberlite XAD-7 resin added to the buffer solution (CH3COONH4, CH3COOH and NH4Cl) of pH = 4.6 ± 0.1. The developed voltammetric method with integrated three-electrode screen-printed sensor was validated with use of certified reference materials (surface, rain and natural water) and can be in future applied to field analyses of Tl(I).