Back to Search
Start Over
Properties of IBA Razor Nano Chamber in small-field radiation therapy using 6 MV FF, 6 MV FFF, and 10 MV FFF photon beams
- Source :
- Acta Oncologica. 60:1419-1424
- Publication Year :
- 2021
- Publisher :
- Informa UK Limited, 2021.
-
Abstract
- Small megavoltage photon fields are increasingly used in modern radiotherapy techniques such as stereotactic radiotherapy. Therefore, it is important to study the reliability of dosimetry in the small-field conditions. The IBA Razor Nano Chamber (Nano chamber) ionization chamber is particularly intended for small-field measurements. In this work, properties of the Nano chamber were studied with both measurements and Monte Carlo (MC) simulations.The measurements and MC simulations were performed with 6 MV, 6 MV FFF and 10 MV FFF photon beams from the Varian TrueBeam linear accelerator. The source-to-surface distance was fixed at 100 cm. The measurements and MC simulations included profiles, percentage depth doses (PDD), and output factors (OF) in square jaw-collimated fields. The MC simulations were performed with the EGSnrc software system in a large water phantom.The measured profiles and PDDs obtained with the Nano chamber were compared against IBA Razor Diode, PTW microDiamond and the PTW Semiflex ionization chamber. These results indicate that the Nano chamber is a high-resolution detector and thus suitable for small field profile measurements down to field sizes 2 × 2 cmThe Nano chamber with its small cavity volume is a potential detector for the small-field dosimetry. In this study, the properties of this detector were characterized with measurements and MC simulations. The determined correction factors
- Subjects :
- Photons
Photon
business.industry
medicine.medical_treatment
Monte Carlo method
Reproducibility of Results
Hematology
General Medicine
Small field
Radiation therapy
Reliability (semiconductor)
Optics
Oncology
Ionization chamber
Nano
medicine
Humans
Dosimetry
Radiology, Nuclear Medicine and imaging
Particle Accelerators
Radiometry
business
Monte Carlo Method
Subjects
Details
- ISSN :
- 1651226X and 0284186X
- Volume :
- 60
- Database :
- OpenAIRE
- Journal :
- Acta Oncologica
- Accession number :
- edsair.doi.dedup.....db12505faa63ec3fb12ca3f681f7e4df
- Full Text :
- https://doi.org/10.1080/0284186x.2021.1979644