An investigation of tomotherapy beam delivery
- Source :
- Medical physics. 24(3)
- Publication Year :
- 1997
-
Abstract
- Experimental simulations for tomotherapy beam delivery were performed using a computer-controlled phantom positioner, a cylindrical phantom, and a 6 MV x-ray slit beam. Both continuous helical beam and sequential segmented tomotherapy (SST) beam deliveries were evaluated. Beam junctioning problem due to couch indexing error or field width errors presented severe dose uniformity perturbations for SST, while the problem was minimized for helical beam delivery. Longitudinal breathing motions were experimentally simulated for helical and SST beam delivery. While motions reduced the dose uniformity perturbations for SST, small artifacts in dose uniformity can be introduced for helical beam delivery. With typical breath frequency and magnitude, for a slit beam of 2.0 cm width at 4 rpm, the dose uniformity perturbation was not significant. A running start/stop technique was implemented with helical beam delivery to sharpen the 20%-80% longitudinal dose fall-off from 1.5 to 0.5 cm. The latter was comparable to the corresponding dose penumbra of a conventional 6 MV 10 x 10 cm2 field. All together, helical beam delivery showed advantages over SST for tomotherapy beam delivery under similar delivery conditions.
- Subjects :
- High energy
medicine.medical_treatment
Physics::Medical Physics
Biophysics
Cylindrical phantom
Tomotherapy
Imaging phantom
Biophysical Phenomena
Dose uniformity
Radiotherapy, High-Energy
Motion
Optics
Beam delivery
medicine
Dosimetry
Humans
Computer Simulation
Technology, Radiologic
Physics
business.industry
Phantoms, Imaging
Radiotherapy Planning, Computer-Assisted
Radiotherapy Dosage
General Medicine
Models, Theoretical
Slit
Evaluation Studies as Topic
Physics::Accelerator Physics
Nuclear medicine
business
Subjects
Details
- ISSN :
- 00942405
- Volume :
- 24
- Issue :
- 3
- Database :
- OpenAIRE
- Journal :
- Medical physics
- Accession number :
- edsair.doi.dedup.....d8dfbc263f4cceac73dc8c0a9ef388ab