A Biomechanical Comparison of High-Tensile Strength Tape Versus High-Tensile Strength Suture for Tendon Fixation Under Cyclic Loading

Authors :
Brad D. Hendershot
Anthony H. Le
Donald R. Fredericks
Lance E. LeClere
Donald F. Colantonio
William B. Roach
Melvin D. Helgeson
Timothy C. Mauntel
Alfred J. Pisano
Sean E. Slaven
Source :
Arthroscopy : the journal of arthroscopicrelated surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association. 37(9)
Publication Year :
2021

Abstract

To compare the biomechanical properties of high-tensile strength tape and high-tensile strength suture across 2 selected stitch techniques, the Krackow and whip stitch, in securing tendinous tissue during 5,000 cycles of nondestructive loading followed by a load to failure.Fourteen matched pairs each of cadaveric Achilles, quadriceps, and patellar tendons (n = 84) were randomly assigned to either Krackow or whip stitch and sutured with either 2-mm high-tensile strength tape or No. 2 high-tensile strength suture. Specimens were preloaded to 20 N, cyclically loaded from 20 to 200 N for 5,000 cycles at 2 Hz, and then loaded to failure at 200 mm/min. Linear mixed models evaluated the effects of suture material and stitch technique on cyclic normalized tendon-suture elongation, total normalized tendon-suture elongation at 5,000 cycles, and maximum load at failure.Across all suture constructs, normalized elongation was greater during the initial 10 cycles, compared with all subsequent cycling intervals (all P.001). There was less total normalized elongation (β = -0.239; P = .007) and greater maximum load at failure in tape (β = 163.71; P = .014) when used in the Krackow stitch compared with the whip stitch.Our findings indicate that tape used in the Krackow stitch maintains the most favorable fixation strength after enduring cyclic loading, with greater maximum load at failure. In addition, overall normalized elongation during long-term cyclic loading was predominately affected by the stitch technique used, regardless of the suture material; however, tape allowed less normalized elongation during the initial loading cycles, especially when placed in the whip stitch.Understanding the potential short- and long-term outcomes of suture material and stitch technique on securing tendinous tissue under repetitive stresses can help inform clinicians on optimal tendon fixation techniques for early postoperative activities.

Details

ISSN :
15263231
Volume :
37
Issue :
9
Database :
OpenAIRE
Journal :
Arthroscopy : the journal of arthroscopicrelated surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association
Accession number :
edsair.doi.dedup.....d8d5d18cc88ce4c29787d398d65ad0ee