Ventilatory parameters measured during a physiological study of simulated powered air-purifying respirator failure in healthy volunteers
- Source :
- Anaesthesia and intensive care. 49(3)
- Publication Year :
- 2021
-
Abstract
- Powered air-purifying respirators (PAPR) are a high level of respiratory personal protective equipment. Like all mechanical devices, they are vulnerable to failure. The precise physiological consequences of failure in live subjects have not previously been reported. We conducted an observational safety study simulating PAPR failure in a group of nine healthy volunteers, wearing loose-fitting hoods, who were observed for a period of ten minutes, or until they requested the experiment be aborted, with continuous monitoring of gas exchange. Relative to baseline, participants demonstrated median reductions in peripheral oxygen saturation of 3.5% (95% confidence interval (CI) –4% to –2%; P = 0.0039) and fraction of inspired oxygen of 0.045 (95% CI –0.05 to –0.04; P = 0.0039), and median increases in inspired partial pressure of carbon dioxide of 27 mmHg (95% CI 23.5–32 mmHg; P = 0.0039), end-tidal partial pressure of carbon dioxide of 11 mmHg (95% CI 7–16 mmHg; P = 0.0039) and minute ventilation of 30 l/min (95% CI 19.4–35.9 l/min; P = 0.0039). Median collateral entrainment of room air into the hood was 17.6 l/min (interquartile range 12.3–27.0 l/min). All subjects reported thermal discomfort, with two (22.2%) requesting early termination of the experiment. Whilst the degree of rebreathing in this experiment was not sufficient to cause dangerous physiological derangement, the degree of reported thermal discomfort combined with the consequences of entrainment of possibly contaminated air into the hood, pose a risk to wearers in the event of failure.
- Subjects :
- Powered air-purifying respirator
business.product_category
business.industry
Carbon Dioxide
Critical Care and Intensive Care Medicine
Confidence interval
Healthy Volunteers
03 medical and health sciences
0302 clinical medicine
Anesthesiology and Pain Medicine
030202 anesthesiology
Interquartile range
Fraction of inspired oxygen
Anesthesia
Room air distribution
Medicine
Humans
030212 general & internal medicine
Respirator
Respiratory system
Respiratory Protective Devices
business
Respiratory minute volume
Subjects
Details
- ISSN :
- 0310057X
- Volume :
- 49
- Issue :
- 3
- Database :
- OpenAIRE
- Journal :
- Anaesthesia and intensive care
- Accession number :
- edsair.doi.dedup.....d6c761018b007c3ba239ae0f8fd63d01