Back to Search Start Over

Faraday-effect polarimeter diagnostic for internal magnetic field fluctuation measurements in DIII-D

Authors :
Weixing Ding
D. Taussig
D. L. Brower
Jie Chen
R. L. Boivin
D. K. Finkenthal
Christopher Muscatello
Source :
The Review of scientific instruments. 87(11)
Publication Year :
2016

Abstract

Motivated by the need to measure fast equilibrium temporal dynamics, non-axisymmetric structures, and core magnetic fluctuations (coherent and broadband), a three-chord Faraday-effect polarimeter-interferometer system with fast time response and high phase resolution has recently been installed on the DIII-D tokamak. A novel detection scheme utilizing two probe beams and two detectors for each chord results in reduced phase noise and increased time response [δb ∼ 1G with up to 3 MHz bandwidth]. First measurement results were obtained during the recent DIII-D experimental campaign. Simultaneous Faraday and density measurements have been successfully demonstrated and high-frequency, up to 100 kHz, Faraday-effect perturbations have been observed. Preliminary comparisons with EFIT are used to validate diagnostic performance. Principle of the diagnostic and first experimental results is presented.

Details

ISSN :
10897623
Volume :
87
Issue :
11
Database :
OpenAIRE
Journal :
The Review of scientific instruments
Accession number :