Back to Search Start Over

Implanted electrodes for multi-month EEG

Authors :
Patrick D. Wolf
Thomas Jochum
Susannah Engdahl
Brad J. Kolls
Source :
EMBC
Publication Year :
2014
Publisher :
IEEE, 2014.

Abstract

An implanted electroencephalogram (EEG) recorder would help diagnose infrequent seizure-like events. A proof-of-concept study quantified the electrical characteristics of the electrodes planned for the proposed recorder. The electrodes were implanted in an ovine model for eight weeks. Electrode impedance was less than 800 Ohms throughout the study. A frequency-domain determination of sedation performed similarly for surface versus implanted electrodes throughout the study. The time-domain correlation between an implanted electrode and a surface electrode was almost as high as between two surface electrodes (0.86 versus 0.92). EEG-certified clinicians judged that the implanted electrode quality was adequate to excellent and that the implanted electrodes provided the same clinical information as surface electrodes except for a noticeable amplitude difference. No significant issues were found that would stop development of the EEG recorder.