A new rapid bedside assay for quantitative testing of D-Dimer (Cardiac D-Dimer) in the diagnostic work-up for deep vein thrombosis

Authors :
Simone Fariselli
Gualtiero Palareti
Michela Cini
Carmela Abate
Cristina Legnani
G. Oca
Source :
Thrombosis Research. 111:149-153
Publication Year :
2003
Publisher :
Elsevier BV, 2003.

Abstract

The accuracy of a new bedside, rapid and quantitative D-Dimer assay (Cardiac D-Dimer) was evaluated in outpatients with clinically suspected deep vein thrombosis (DVT); VIDAS test was used as reference method. Eighty consecutive outpatients with suspected DVT of a lower limb were included in the study. Patients were classified as DVT positive or negative according to results of objective test (serial CUS), pretest clinical probability and 3-month follow-up. DVT was diagnosed in 32/80 patients (40%). The performance of the two D-Dimer assays was comparable, as indicated by the areas under the ROC curves (0.89 and 0.88, for Cardiac D-Dimer and VIDAS, respectively) and the coefficient of correlation (r=0.91). The reproducibility of the test was acceptable (from 6.2% to 12.0%). The sensitivity and negative predictive values were 100% for both tests. The specificity (SP) and positive predictive values (PPV) were similar (SP: 50.0% and 52.0%, PPV: 57.1% and 58.2%, for Cardiac D-Dimer and VIDAS, respectively). The Cardiac D-Dimer test proved to be very accurate and produced results fully comparable to those obtained with the VIDAS test. Since the test can be directly performed in the emergency room within a few minutes, it seems to have great clinical potential. The place of this assay in the diagnostic strategy of DVT remains to be determined in prospective management studies.

Details

ISSN :
00493848
Volume :
111
Database :
OpenAIRE
Journal :
Thrombosis Research
Accession number :
edsair.doi.dedup.....d4b809353acfd1e994d17c488cab7c37
Full Text :
https://doi.org/10.1016/j.thromres.2003.08.027