Validation of three automatic devices for self-measurement of blood pressure according to the International Protocol: The Omron M3 Intellisense (HEM-7051-E), the Omron M2 Compact (HEM 7102-E), and the Omron R3-I Plus (HEM 6022-E)
- Source :
- Blood Pressure Monitoring. 15:49-54
- Publication Year :
- 2010
- Publisher :
- Ovid Technologies (Wolters Kluwer Health), 2010.
-
Abstract
- Three oscillometric devices for the self blood pressure measurement were evaluated according to the International Protocol of the European Society of Hypertension in three separate studies. The Omron M3 Intellisense and the Omron M2 Compact measures blood pressure (BP) at the brachial level; the Omron R3-I Plus measures BP at the wrist level.The International Protocol is divided into two phases and includes a total number of 33 participants on whom the validation is performed. In each study, and for each participant, four BP measurements were taken simultaneously by two observers using mercury sphygmomanometers alternately with three measurements with the tested device. The difference between the observers and the device BP values was calculated for each measurement. The 99 pairs of BP differences were classified into three categories (or =5,or =10,or =15 mmHg).All three tested devices passed the first and second phase of the validation process. The mean differences between the device and mercury readings were 1.9+/-3.0 and -1.0+/-2.3 mmHg for systolic and diastolic BP, respectively, for the Omron M3 Intellisense device, 2.5+/-5.4 and -2.3+/-3.6 mmHg for the Omron M2 Compact device, and 1.4+/-4.5 and 0.8+/-4.6 mmHg for the Omron R3-I Plus device.Readings of the Omron M3 Intellisense, the Omron M2 Compact and the Omron R3-I Plus, differing by less than 5, 10, and 15 mmHg fulfill the International Protocol requirements and therefore can be used by patients for self blood pressure measurement.
- Subjects :
- Adult
Male
medicine.medical_specialty
Validation study
Assessment and Diagnosis
Self measurement
Internal Medicine
Humans
Medicine
Hypertension diagnosis
Aged
Advanced and Specialized Nursing
Protocol (science)
business.industry
Reproducibility of Results
Blood Pressure Determination
General Medicine
Middle Aged
Sphygmomanometers
Self Care
Blood pressure
Hypertension
Physical therapy
Female
Cardiology and Cardiovascular Medicine
business
Subjects
Details
- ISSN :
- 13595237
- Volume :
- 15
- Database :
- OpenAIRE
- Journal :
- Blood Pressure Monitoring
- Accession number :
- edsair.doi.dedup.....d309ae77bab0943471f44b3c289445eb
- Full Text :
- https://doi.org/10.1097/mbp.0b013e3283354b11