To clarify the availability of the dielectrophoretic impedance measurement (DEPIM) system as the evaluator for oral care, we evaluated the usefulness of DEPIM system by comparison with the standard plate counting (SPC) method. First, the relationship between the DEPIM results and bacterial concentration measured by SPC was clarified. Next, the measurement of the microorganism number in a mixed suspension was evaluated with DEPIM and SPC. The bacterial counts with DEPIM strongly correlated with those with SPC (r(2)=0.633-0.997) and this correlation was also shown in the measurement of a mixed bacterial suspension (ranging from 10(5) to 10(8) cfu/ml) of two bacterial species. Moreover, the experiments using dissociating enzymes to eliminate the influence of the size of the bacterial aggregates demonstrated that the microbial measurement results with DEPIM are unaffected by bacterial aggregates. This study demonstrated that bacterial counts with DEPIM strongly correlated with those with SPC and were unaffected by bacterial aggregates.