Monte Carlo simulated corrections for beam commissioning measurements with circular and MLC shaped fields on the CyberKnife M6 System: a study including diode, microchamber, point scintillator, and synthetic microdiamond detectors

Authors :
J M Noll
N Satariano
Serenella Russo
Paolo Francescon
W. Kilby
Laura Masi
Source :
Physics in Medicine and Biology. 62:1076-1095
Publication Year :
2017
Publisher :
IOP Publishing, 2017.

Abstract

Monte Carlo simulation was used to calculate correction factors for output factor (OF), percentage depth-dose (PDD), and off-axis ratio (OAR) measurements with the CyberKnife M6 System. These include the first such data for the InCise MLC. Simulated detectors include diodes, air-filled microchambers, a synthetic microdiamond detector, and point scintillator. Individual perturbation factors were also evaluated. OF corrections show similar trends to previous studies. With a 5 mm fixed collimator the diode correction to convert a measured OF to the corresponding point dose ratio varies between −6.1% and −3.5% for the diode models evaluated, while in a 7.6 mm × 7.7 mm MLC field these are −4.5% to −1.8%. The corresponding microchamber corrections are +9.9% to +10.7% and +3.5% to +4.0%. The microdiamond corrections have a maximum of −1.4% for the 7.5 mm and 10 mm collimators. The scintillator corrections are 15%, reducing to d max were