Back to Search
Start Over
Type of screw does not influence holding power in the femoral head: A cadaver study with shearing test
- Source :
- Acta Orthopaedica Scandinavica. 70:435-438
- Publication Year :
- 1999
- Publisher :
- Medical Journals Sweden AB, 1999.
-
Abstract
- We studied the holding power during cut-out tests of 4 different screw fixation devices of the femoral head, correlated to head size and bone density, on 48 paired deep frozen proximal femurs of fresh gross specimens. We found no significant effect of the device on the holding power. There was a significantly higher maximal load at failure in large and dense heads. It seems then unnecessary to advocate more sophisticated but also more expensive devices to fix femoral neck fractures.
- Subjects :
- Male
Compressive Strength
Friction
Bone Screws
Prosthesis Design
Femoral Neck Fractures
Weight-Bearing
Femoral head
Bone Density
Cadaver
Tensile Strength
Materials Testing
medicine
Humans
Orthopedics and Sports Medicine
Femur
Aged
Femoral neck
Fixation (histology)
Aged, 80 and over
Orthodontics
Osteosynthesis
business.industry
Biomechanics
Femur Head
Anatomy
Middle Aged
Prosthesis Failure
Equipment Failure Analysis
medicine.anatomical_structure
Female
Surgery
business
Subjects
Details
- ISSN :
- 00016470
- Volume :
- 70
- Database :
- OpenAIRE
- Journal :
- Acta Orthopaedica Scandinavica
- Accession number :
- edsair.doi.dedup.....ce25c2a7f2488f6452875134534103bb
- Full Text :
- https://doi.org/10.3109/17453679909000977