Attenuation correction for small animal PET tomographs

Authors :
Patrick L Chow
Fernando R. Rannou
Arion F. Chatziioannou
Source :
Chow, Patrick L; Rannou, Fernando R; & Chatziioannou, Arion F. (2005). Attenuation correction for small animal PET tomographs. Physics in Medicine and Biology, 50(8), 1837-1850. UCLA: Retrieved from: http://www.escholarship.org/uc/item/41n377p3
Publication Year :
2005

Abstract

Attenuation correction is one of the important corrections required for quantitative positron emission tomography (PET). This work will compare the quantitative accuracy of attenuation correction using a simple global scale factor with traditional transmission-based methods acquired either with a small animal PET or a small animal x-ray computed tomography (CT) scanner. Two phantoms (one mouse-sized and one rat-sized) and two animal subjects (one mouse and one rat) were scanned in CTI Concorde Microsystem's MicroPET(R) Focus(TM) for emission and transmission data and in ImTek's MicroCAT(TM) II for transmission data. PET emission image values were calibrated against a scintillation well counter. Results indicate that the scale factor method of attenuation correction places the average measured activity concentration about the expected value, without correcting for the cupping artefact from attenuation. Noise analysis in the phantom studies with the PET-based method shows that noise in the transmission data increases the noise in the corrected emission data. The CT-based method was accurate and delivered low-noise images suitable for both PET data correction and PET tracer localization.

Details

ISSN :
00319155
Volume :
50
Issue :
8
Database :
OpenAIRE
Journal :
Physics in medicine and biology
Accession number :
edsair.doi.dedup.....cb5b7ffd8eb7db775c5523df051b69b0