Back to Search Start Over

Adaptive optics photoacoustic microscopy

Authors :
Hao Zhang
Xiangyang Zhang
Carmen A. Puliafito
Minshan Jiang
Shuliang Jiao
Source :
Optics Express
Publication Year :
2010
Publisher :
The Optical Society, 2010.

Abstract

We have developed an adaptive optics photoacoustic microscope (AO-PAM) for high-resolution imaging of biological tissues, especially the retina. To demonstrate the feasibility of AO-PAM we first designed the AO system to correct the wavefront errors of the illuminating light of PAM. The aberrations of the optical system delivering the illuminating light to the sample in PAM was corrected with a close-loop AO system consisting of a 141-element MEMS-based deformable mirror (DM) and a Shack-Hartmann (SH) wavefront sensor operating at 15 Hz. The photoacoustic signal induced by the illuminating laser beam was detected by a custom-built needle ultrasonic transducer. When the wavefront errors were corrected by the AO system, the lateral resolution of PAM was measured to be better than 2.5 µm using a low NA objective lens. We tested the system on imaging ex vivo ocular samples, e.g., the ciliary body and retinal pigment epithelium (RPE) of a pig eye. The AO-PAM images showed significant quality improvement. For the first time we were able to resolve single RPE cells with PAM.

Details

ISSN :
10944087
Volume :
18
Database :
OpenAIRE
Journal :
Optics Express
Accession number :
edsair.doi.dedup.....cb12fe283abbf8735b96f8aafade57ce
Full Text :
https://doi.org/10.1364/oe.18.021770