Back to Search Start Over

Time and position sensitive single photon detector for scintillator read-out

Authors :
K. Tittelmeier
Volker Dangendorf
M. Brandis
S. Schössler
B. Bromberger
Achim Czasch
L. Ph. H. Schmidt
Ottmar Jagutzki
Source :
Journal of Instrumentation. 7:C02048-C02048
Publication Year :
2012
Publisher :
IOP Publishing, 2012.

Abstract

We have developed a photon counting detector system for combined neutron and gamma radiography which can determine position, time and intensity of a secondary photon flash created by a high-energy particle or photon within a scintillator screen. The system is based on a micro-channel plate photomultiplier concept utilizing image charge coupling to a position- and time-sensitive read-out anode placed outside the vacuum tube in air, aided by a standard photomultiplier and very fast pulse-height analyzing electronics. Due to the low dead time of all system components it can cope with the high throughput demands of a proposed combined fast neutron and dual discrete energy gamma radiography method (FNDDER). We show tests with different types of delay-line read-out anodes and present a novel pulse-height-to-time converter circuit with its potential to discriminate gamma energies for the projected FNDDER devices for an automated cargo container inspection system (ACCIS).<br />Proceedings of FNDA 2011

Details

ISSN :
17480221
Volume :
7
Database :
OpenAIRE
Journal :
Journal of Instrumentation
Accession number :
edsair.doi.dedup.....caa56ab9a54aa84b3e1fdec9e5d70780
Full Text :
https://doi.org/10.1088/1748-0221/7/02/c02048