Back to Search Start Over

Multipoint heterodyne vibrometry for measurements on unstable biomedical objects

Authors :
Joris J.J. Dirckx
H.J. van Elburg
Willem F. Decraemer
Source :
Proceedings of the Society of Photo-optical Instrumentation Engineers
Publication Year :
2005
Publisher :
SPIE, 2005.

Abstract

In unstable objects, such as biological specimens, vibrational patterns can change quickly with time. To determine phase relationships between several vibrating points in such an object, it is therefore necessary to measure those points simultaneously. We describe a dual channel heterodyne interferometer capable to acquire vibrational data at two arbitrary spatial points simultaneouly over the entire acoustic frequency range. The design allows extension to four or even eight independent channels. Channel cross-talk is measured to be about -30 dB, below the S/N ratio within the signal channel. Velocity resolution is estimated to be 10 mu m/s, corresponding to an amplitude resolution of 1.5 nm at 1 kHz.

Details

ISSN :
0277786X
Database :
OpenAIRE
Journal :
Eighth International Symposium on Laser Metrology
Accession number :
edsair.doi.dedup.....c97306cafbf2136968f8dafd65bd1976
Full Text :
https://doi.org/10.1117/12.611631