Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT
- Source :
- EJNMMI Physics, EJNMMI Physics, Vol 6, Iss 1, Pp 1-17 (2019)
- Publication Year :
- 2019
- Publisher :
- Springer International Publishing, 2019.
-
Abstract
- Background This paper describes the National Electrical Manufacturers Association (NEMA) system performance of the Discovery MI 3-ring PET/CT (GE Healthcare) installed in Bruges, Belgium. This time-of-flight (TOF) PET camera is based on silicon photomultipliers instead of photomultiplier tubes. Methods The NEMA NU2-2012 standard was used to evaluate spatial resolution, sensitivity, image quality (IQ) and count rate curves of the system. Timing and energy resolution were determined. Results Full width at half maximum (FWHM) of spatial resolution in radial, tangential and axial direction was 4.69, 4.08 and 4.68 mm at 1 cm; 5.58, 4.64 and 5.83 mm at 10 cm; and 7.53, 5.08 and 5.47 mm at 20 cm from the centre of the field of view (FOV) for the filtered backprojection reconstruction. For non-TOF ordered subset expectation maximization (OSEM) reconstruction without point spread function (PSF) correction, FWHM was 3.87, 3.69 and 4.15 mm at 1 cm; 4.80, 3.81 and 4.87 mm at 10 cm; and 7.38, 4.16 and 3.98 mm at 20 cm. Sensitivity was 7.258 cps/kBq at the centre of the FOV and 7.117 cps/kBq at 10-cm radial offset. Contrast recovery (CR) using the IQ phantom for the TOF OSEM reconstruction without PSF correction was 47.4, 59.3, 67.0 and 77.0% for the 10-, 13-, 17- and 22-mm radioactive spheres and 82.5 and 85.1% for the 28- and 37-mm non-radioactive spheres. Background variability (BV) was 16.4, 12.1, 9.1, 6.6, 5.1 and 3.8% for the 10-, 13-, 17-, 22-, 28- and 37-mm spheres. Lung error was 8.5%. Peak noise equivalent count rate (NECR) was 102.3 kcps at 23.0 kBq/ml with a scatter fraction of 41.2%. Maximum accuracy error was 3.88%. Coincidence timing resolution was 375.6 ps FWHM. Energy resolution was 9.3% FWHM. Q.Clear reconstruction significantly improved CR and reduced BV compared with OSEM. Conclusion System sensitivity and NECR are lower and IQ phantom’s BV is higher compared with larger axial FOV (AFOV) scanners like the 4-ring discovery MI, as expected from the smaller solid angle of the 3-ring system. The other NEMA performance parameters are all comparable with those of the larger AFOV scanners.
- Subjects :
- lcsh:Medical physics. Medical radiology. Nuclear medicine
Point spread function
Photomultiplier
Image quality
lcsh:R895-920
Biomedical Engineering
Imaging phantom
030218 nuclear medicine & medical imaging
03 medical and health sciences
0302 clinical medicine
Silicon photomultiplier
Optics
Ordered subset expectation maximization
Silicon photomultiplier based PET/CT
Radiology, Nuclear Medicine and imaging
Time-of-flight PET/CT
Instrumentation
Image resolution
Original Research
Physics
Radiation
NEMA
business.industry
Full width at half maximum
030220 oncology & carcinogenesis
business
Discovery MI 3-ring
Subjects
Details
- Language :
- English
- ISSN :
- 21977364
- Volume :
- 6
- Database :
- OpenAIRE
- Journal :
- EJNMMI Physics
- Accession number :
- edsair.doi.dedup.....c95ff43a357ace433a58f0ee32bd2661