The Impact on Drug Mass Flow Rate of Interrupting and Resuming Carrier Fluid Flow

Authors :
Bertrand Debaene
Pascal Odou
Damien Lannoy
Bertrand Décaudin
Nicolas Simon
Christine Barthélémy
Source :
Anesthesia & Analgesia. 114:328-332
Publication Year :
2012
Publisher :
Ovid Technologies (Wolters Kluwer Health), 2012.

Abstract

Stopping and resuming carrier fluid flow can lead to potentially dangerous transient disturbances in drug mass flow rate. We compared the impact of 2 infusion sets, one with very low dead-space volume and the other with greater dead-space volume, on the amount of drug delivered during stop-and-go carrier fluid flows.Two infusion sets, both with antireflux, connected to an angiocatheter and with dead-space volumes of 6.185 mL and 0.071 mL, respectively, were assessed. Two protocols were studied: carrier fluid flow of 90 mL/h associated with noradrenaline infused at 7 mL/h and carrier fluid flow of 350 mL/h with a noradrenaline infusion flow of 65 mL/h. During both protocols, the carrier fluid was stopped and resumed at the same rate 30 minutes later. Effluent noradrenaline concentration was measured using UV spectrophotometry. Flow change efficiency was calculated from the ratio of the area under the experimental mass flow rate curve to the area under the theoretical instantaneous mass flow rate curve.For both flow rate conditions, flow change efficiency was significantly different for the 2 infusion sets during the 10-minute period after stopping carrier fluid flow and the 10-minute period after restarting it. The major phenomena were sudden decreases in drug delivery after stopping carrier flow and sudden, temporary increases when it was resumed. The very low dead-space volume infusion set resulted in significant reduction in changes in drug delivery compared with the standard set, even at high flow rates.The use of a very low dead-space volume set attenuates disturbances in drug delivery caused by interrupting and resuming carrier fluid flow.

Details

ISSN :
00032999
Volume :
114
Database :
OpenAIRE
Journal :
Anesthesia & Analgesia
Accession number :
edsair.doi.dedup.....c862e8ba81bef841863986e857942736