Technical Note: Use of automation to eliminate shift errors

Authors :
Elizabeth L. Covington
Richard A. Popple
Rex A. Cardan
Source :
Journal of Applied Clinical Medical Physics
Publication Year :
2020
Publisher :
Wiley, 2020.

Abstract

Purpose To create automated tools within the treatment planning system (TPS) that eliminate the common error pathway of providing incorrect shift instructions to therapists. Materials/Methods Two scripts were created within the TPS using the Eclipse API (Varian Medical Systems, Palo Alto, CA). One script detects whether or not the user origin has been placed correctly at the intersection of the simulation markers while the other calculates a shift instruction sheet that can be printed for treatment. Results Analysis of our RO‐ILS database identified eight errors caused by improper setting of the user origin in the treatment planning system. The user origin script flagged all of the treatment plans for markers inconsistent with user origin. Automated calculation of shifts eliminated the error pathway of miscalculating or transcribing shift values. Conclusion Automation can eliminate the common error pathway of providing the wrong shifts to therapists. The scripts have been made available as open‐source software for implementation at other radiotherapy clinics.