Abstract
- Purpose To create automated tools within the treatment planning system (TPS) that eliminate the common error pathway of providing incorrect shift instructions to therapists. Materials/Methods Two scripts were created within the TPS using the Eclipse API (Varian Medical Systems, Palo Alto, CA). One script detects whether or not the user origin has been placed correctly at the intersection of the simulation markers while the other calculates a shift instruction sheet that can be printed for treatment. Results Analysis of our RO‐ILS database identified eight errors caused by improper setting of the user origin in the treatment planning system. The user origin script flagged all of the treatment plans for markers inconsistent with user origin. Automated calculation of shifts eliminated the error pathway of miscalculating or transcribing shift values. Conclusion Automation can eliminate the common error pathway of providing the wrong shifts to therapists. The scripts have been made available as open‐source software for implementation at other radiotherapy clinics.
- Subjects :
- Organs at Risk
shift errors
Common error
Computer science
Radiotherapy Setup Errors
computer.software_genre
Automation
Software
Neoplasms
Technical Note
scripting
Humans
Radiology, Nuclear Medicine and imaging
Instrumentation
Retrospective Studies
Eclipse
Medical systems
Radiation
Intersection (set theory)
business.industry
Radiotherapy Planning, Computer-Assisted
Radiotherapy Dosage
Technical note
RO‐ILS
Scripting language
Radiotherapy, Intensity-Modulated
Technical Notes
business
Software engineering
computer
Subjects
Details
- ISSN :
- 15269914
- Volume :
- 21
- Database :
- OpenAIRE
- Journal :
- Journal of Applied Clinical Medical Physics
- Accession number :