Back to Search
Start Over
Clinical Accuracy and Safety Using the SynchroMed II Intrathecal Drug Infusion Pump
- Source :
- Regional Anesthesia and Pain Medicine
- Publication Year :
- 2014
- Publisher :
- Lippincott Williams & Wilkins, 2014.
-
Abstract
- Background and Objectives We evaluated the infusion accuracy and device-related safety of implantable drug infusion pumps in subjects with chronic pain or severe spasticity. Methods Nine centers in the United States enrolled patients receiving intrathecal drug delivery systems to manage chronic pain and/or severe spasticity. Infusion accuracy was assessed at 6 and 12 months by comparing syringe-measured delivered volumes to programmer-predicted volumes. Safety was evaluated through analysis of adverse events. Separate laboratory testing conducted by the manufacturer also evaluated infusion accuracy. Results Eighty of 82 enrolled subjects were implanted. Sixty-five and 54 subjects, respectively, were analyzable for accuracy at 6 and 12 months. On average at 6 months, the pumps were measured to have delivered 1% more than the programmed delivery volume. Analyzed on a per-refill basis, the pumps delivered, on average, 2.5% more than the programmed delivery volume. Differences between per-refill means versus per-subject cumulative means were due to limitations in clinicians’ ability to precisely visualize single small syringe-volume differences, or possibly incomplete withdrawal of fluid from the pump. Laboratory testing demonstrated a per-refill mean accuracy error of minus 2.4%. Because average observed flow-rate error at 6 and 12 months (1% overinfusion) was derived from pump residual volume measurements by syringe and carried out in a clinical setting, clinical volume ratios were larger than direct volume measurements by weight observed in the laboratory. No deaths, permanent injuries, or unanticipated adverse device effects occurred. Conclusions The pump accurately delivered intrathecal medication in the clinical setting of this study. Adverse events were similar in nature and severity to those described in the product labeling and literature.
- Subjects :
- Adult
Male
medicine.medical_specialty
Overinfusion
Endpoint Determination
Pain
Intrathecal
Patient safety
Young Adult
Medicine
Humans
Spasticity
Adverse effect
Syringe
Infusion Pumps
Injections, Spinal
Aged
Aged, 80 and over
business.industry
Chronic pain
Reproducibility of Results
Drug infusion
General Medicine
Original Articles
Middle Aged
medicine.disease
Surgery
Anesthesiology and Pain Medicine
Muscle Spasticity
Anesthesia
Female
Patient Safety
medicine.symptom
business
Subjects
Details
- Language :
- English
- ISSN :
- 15328651 and 10987339
- Volume :
- 39
- Issue :
- 4
- Database :
- OpenAIRE
- Journal :
- Regional Anesthesia and Pain Medicine
- Accession number :
- edsair.doi.dedup.....c803e3d90a2e884a78bb5b17c39c99a7