Back to Search
Start Over
Automated IR determination of petroleum products in water based on sequential injection analysis
- Source :
- Talanta. 148
- Publication Year :
- 2015
-
Abstract
- The simple and easy performed automated method for the IR determination of petroleum products (PP) in water using extraction-chromatographic cartridges has been developed. The method assumes two stages: on-site extraction of PP during a sampling by using extraction-chromatographic cartridges and subsequent determination of the extracted PP using sequential injection analysis (SIA) with IR detection. The appropriate experimental conditions for extraction of the dissolved in water PP and for automated SIA procedure were investigated. The calibration plot constructed using the developed procedure was linear in the range of 3–200 μg L −1 . The limit of detection (LOD), calculated from a blank test based on 3σ was 1 µg L −1 . The sample volume was 1 L. The system throughput was found to be 12 h −1 .
- Subjects :
- Detection limit
Sequential injection analysis
business.industry
Chemistry
Calibration curve
010401 analytical chemistry
Extraction (chemistry)
Analytical chemistry
Sampling (statistics)
010501 environmental sciences
01 natural sciences
Water based
0104 chemical sciences
Analytical Chemistry
Cartridge
Petroleum product
business
0105 earth and related environmental sciences
Subjects
Details
- ISSN :
- 18733573
- Volume :
- 148
- Database :
- OpenAIRE
- Journal :
- Talanta
- Accession number :