Back to Search Start Over

Helical differential X-ray phase-contrast computed tomography

Authors :
Danays Kunka
Jian Fu
Renbo Tan
Julia Herzen
Juergen Mohr
Marian Willner
Martin Bech
Liyuan Chen
Klaus Achterhold
Franz Pfeiffer
Source :
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 30(3)
Publication Year :
2013

Abstract

We report on the first experimental results of helical differential phase-contrast computed tomography (helical DPC-CT) with a laboratory X-ray tube source and a Talbot-Lau grating interferometer. The results experimentally verify the feasibility of helical data acquisition and reconstruction in phase-contrast imaging, in analogy to its use in clinical CT systems. This allows fast and continuous volumetric scans for long objects with lengths exceeding the dimension of the detector. Since helical CT revolutionized the field of medical CT several years ago, we anticipate that this method will bring the same significant impact on the future medical and industrial applications of X-ray DPC-CT.

Details

ISSN :
1724191X
Volume :
30
Issue :
3
Database :
OpenAIRE
Journal :
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)
Accession number :
edsair.doi.dedup.....c1ba67aca069441973c0ca619439df9e