Noncontact Evaluation of Corneal Grafts: Swept-Source Fourier Domain OCT Versus High-Resolution Scheimpflug Imaging
- Source :
- Cornea. 36(4)
- Publication Year :
- 2017
-
Abstract
- To study the reliability and repeatability of anterior segment measurements in patients who underwent penetrating keratoplasty with swept light source Fourier domain anterior segment optical coherence tomography (SS-OCT) in comparison with a high-resolution Scheimpflug camera.Keratometry (K) and elevation data of the anterior and posterior corneal surfaces, central corneal thickness, and anterior chamber depth were determined in 56 postkeratoplasty eyes (mean age: 59.6 ± 16.5 yrs) with SS-OCT (CASIA SS-1000; Tomey) followed by Scheimpflug photography (Pentacam High Resolution; Oculus). Examinations were also performed on 48 right eyes of 48 age-matched healthy subjects (mean age: 66.3 ± 7.0 yrs).A statistically significant difference was detected between both tomographers in anterior flat K [1.7 ± 4.92 diopters (D)], posterior astigmatism (1.32 ± 5.7 D) and elevation (6.65 ± 14.07 μm), and central corneal thickness (27.39 ± 90.81 μm) and anterior chamber depth values (0.14 ± 0.66 mm) in patients who had undergone corneal transplantation (P0.05). In the postkeratoplasty group, repeatability of SS-OCT was higher for every anterior segment parameter in comparison with the Pentacam. In corneal grafts, the variability between 3 consecutive measurements on the same subject was remarkably lower for every parameter while using SS-OCT.In patients who had undergone corneal transplantation, SS-OCT provided more reliable and repeatable anterior segment evaluations when compared with the rotating Scheimpflug system. Our findings suggested that the tested high-resolution swept-source OCT system is a comprehensive and precise diagnostic tool for postoperative follow-up of patients who have undergone penetrating keratoplasty.
- Subjects :
- Male
Materials science
genetic structures
Corneal Pachymetry
Scheimpflug principle
Klinikai orvostudományok
01 natural sciences
law.invention
Corneal Diseases
010309 optics
Cornea
03 medical and health sciences
symbols.namesake
0302 clinical medicine
Optical coherence tomography
law
Anterior Eye Segment
0103 physical sciences
medicine
Photography
Humans
Corneal pachymetry
Fourier domain
Aged
medicine.diagnostic_test
Keratometer
Fourier Analysis
Reproducibility of Results
Orvostudományok
Repeatability
Middle Aged
eye diseases
Healthy Volunteers
Ophthalmology
Fourier analysis
030221 ophthalmology & optometry
symbols
Female
sense organs
Tomography
Keratoplasty, Penetrating
Tomography, Optical Coherence
Biomedical engineering
Subjects
Details
- ISSN :
- 15364798
- Volume :
- 36
- Issue :
- 4
- Database :
- OpenAIRE
- Journal :
- Cornea
- Accession number :
- edsair.doi.dedup.....bf398e2661079192d616654cbe6e298a