Comparison of Sonography versus Digital Breast Tomosynthesis to Locate Intramammary Marker Clips

Authors :
Christian R. Loehberg
Ruediger Schulz-Wendtland
G Dilbat
Sebastian M. Jud
Katharina Heusinger
Peter Dankerl
Michael Uder
Michael P. Lux
P. A. Fasching
Mayada R. Bani
Claudia Rauh
M. Meier-Meitinger
B. Brehm
Christian M. Bayer
M. W. Beckmann
David L. Wachter
Source :
Geburtshilfe und Frauenheilkunde. 75:72-76
Publication Year :
2015
Publisher :
Georg Thieme Verlag KG, 2015.

Abstract

Introduction: This study aimed to compare the accuracy of sonography versus digital breast tomosynthesis to locate intramammary marker clips placed under ultrasound guidance. Patients and Methods: Fifty patients with suspicion of breast cancer (lesion diameter less than 2 cm [cT1]) had ultrasound-guided core needle biopsy with placement of a marker clip in the center of the tumor. Intramammary marker clips were subsequently located with both sonography and digital breast tomosynthesis. Results: Sonography detected no dislocation of intrammammary marker clips in 42 of 50 patients (84 %); dislocation was reported in 8 patients (16 %) with a maximum dislocation of 7 mm along the x-, y- or z-axis. Digital breast tomosynthesis showed accurate placement without dislocation of the intramammary marker clip in 48 patients (96 %); 2 patients (4 %) had a maximum clip dislocation of 3 mm along the x-, y- or z-axis (p

Details

ISSN :
14388804 and 00165751
Volume :
75
Database :
OpenAIRE
Journal :
Geburtshilfe und Frauenheilkunde
Accession number :
edsair.doi.dedup.....beebe270b758bc1e93b2c0b06adea679
Full Text :
https://doi.org/10.1055/s-0034-1396164