Evaluation of the BacT/Alert® microbial detection system with FAN aerobic and FAN anaerobic bottles for culturing normally sterile body fluids other than blood

Authors :
Marie Louie
Karen Scythes
Andrew E. Simor
Helen Meaney
Source :
Diagnostic Microbiology and Infectious Disease. 37:5-9
Publication Year :
2000
Publisher :
Elsevier BV, 2000.

Abstract

We evaluated the BacT/Alert Microbial Detection System (Organon Teknika Corporation, Durham, NC, USA) by using FAN bottles compared to conventional culture methods for the recovery of microorganisms from normally sterile body fluids other than blood and dialysates. Clinically significant pathogens were isolated from 116 (11%) of 1, 099 consecutive specimens (80 from both conventional media and FAN bottles; 23 from FAN bottles only; 13 from conventional media only). Gram-positive cocci were more likely to be recovered from FAN bottles than from conventional media (p = 0.04). Contaminants were also more likely to have grown in FAN bottles (3%) than on conventional media (1%) (p = 0.04). The mean time to detection of significant pathogens was 20.9 h using FAN bottles as compared to 30. 9 h using conventional media (p = 0.0001). These results indicate that the BacT/Alert Microbial Detection System using FAN blood culture bottles improves the yield of clinically significant Gram-positive isolates from normally sterile body fluids with a reduced time to detection.

Details

ISSN :
07328893
Volume :
37
Database :
OpenAIRE
Journal :
Diagnostic Microbiology and Infectious Disease
Accession number :
edsair.doi.dedup.....bdbb8ca89685e654a3f0f104f06d1075
Full Text :
https://doi.org/10.1016/s0732-8893(99)00157-1