Back to Search Start Over

An automated system to define the optimal operating settings of cryogenic calorimeters

Authors :
S. Di Domizio
Irene Nutini
L. Canonica
L. Marini
G. Pessina
Paolo Carniti
Andrea Giachero
K. Alfonso
Claudio Gotti
C. Bucci
Alfonso, K
Bucci, C
Canonica, L
Carniti, P
Di Domizio, S
Giachero, A
Gotti, C
Marini, L
Nutini, I
Pessina, G
Publication Year :
2021
Publisher :
Elsevier B.V., 2021.

Abstract

Cryogenic macro-calorimeters instrumented with NTD thermistors have been developed for several decades. The choice of the optimal bias current is crucial for a proper operation of these detectors, both in terms of energy resolution and stability. In this paper we present a set of automatic measurements and analysis procedures for the characterization and optimization of the working configuration of the NTD thermistors. The presented procedures were developed for CUORE , an array of 988 cryogenic macro-calorimeters instrumented with NTD thermistors that has been taking data since 2017. These procedures made it possible to characterize a large number of detectors in a reliable way. They are suitable enough to be used also in other large arrays of cryogenic detectors, such as CUPID.

Details

Language :
English
Database :
OpenAIRE
Accession number :
edsair.doi.dedup.....bcc3e328b054145bbd534616c9f30f81